7.
Press the button Start
8.
Depending on the position of the test current
clockwise completely rotated) the following message will turn up:
9.
Turning the test current
Adjust this control until obtaining the desired value or until the current
indicator (bar-graph) indicates 100%. The bargraph will show the test
current value as a percentage of the nominal value selected.
10. The lowest current for measuring is 10% of the range current nominal
value. It is important to consider that the measurement errors
increase while test current decreases. The error is specified for test
currents higher than 80%.
11. The display
12. The corresponding unit will be expressed as Ω (ohms) or mΩ (milli-
ohms).
13. The measured value can be retained in the display
the hold
key. Pressing this key again, the value will be updated.
14. Press the key Stop
turn the equipment OFF using the On/Off
the equipment before pressing the key Stop
PRECAUTION: Do not connect or disconnect the test leads during the
measurement.
.
LOW CURRENT
control clockwise increases the current.
will show the resistance value measured.
in order to finish the measurement, and then
12
control (counter-
00:01
by pressing
switch. (Do not turn off
).