INCUBATOR COMPONENTS
24. Access port: Place the silicon caps on both the outside and inside of the port when the port is not
being used.
25. Remote alarm terminal: Sends the alarm to a remote location by connecting to an external alarm unit.
Refer to page 17.
LCD touch panel
The following display (called the Top screen) will appear when the power switch is turned ON.
Note: It takes approximately 20 seconds until the Top screen is displayed. During warming-up, "Status:
Gas sensor initializing" is displayed in the Message display field (18), and "--.-" is displayed in the present
CO
density display field (5) and the present O
2
3
1
1. Present temperature display field
The current chamber temperature is displayed.
2. Set temperature value display field
The set value of chamber temperature is
displayed. Default setting: 37
14
density display field (10).
2
2
o
C
3. Heating indicator
This lights up when the heater is on.
4. Present date/time display field
The date and time is set when the incubator is
shipped from the factory. Refer to page 51 for
details.
4