(Table 2)
MDSTH1002 IR Thermometer is intended for use in the electromagnetic
environment specified below. The customer or users of this Clinical IR
Thermometers should assure that it is used in such environments.
Immunity test
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC61000-4-4
Surge
IEC 61000-4-5
Voltage dips
short
interruptions
and voltage
variations on
power supply
IEC 61000-4-11
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
24 MDSTH1002 (PIC-00364) | Revision: 09/17/20
Guidance and manufacturer's declaration
—electromagnetic immunity
IEC 60601
Compliance
Test level
± 6 kV
± 6 kV
contact ±8
contact ±8
kV air
kV air
Not
Not
applicable
applicable
Not
Not
applicable
applicable
Not
Not
applicable
applicable
3A/m
3A/m
Electromagnetic
level
environment – guidance
Floor should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Not applicable
Not applicable
Not applicable
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.