(Table 4)
Guidance and manufacturer's declaration
MDSTH1002 IR Thermometer is intended for use in the electromagnetic
environment specified below. The customer or users of this Clinical IR
Thermometers should assure that it is used in such environments.
Immunity
IEC 60601
test
Test level
Conducted
RF
Not
IEC 61000
applicable
-4-6
Radiated
3 V/m
RF IEC
80 MHz to
61000-4- 3
2.5 GHz
—electromagnetic immunity
Compliance
level
Portable and mobile RF
communications equipment
should be used no closer to
any part of the MDSTH1002
IR Thermometer, including
cables, than the recommended
separation distance calculated
from the equation appropriate to
the frequency of the transmitter.
Recommended separation distance
d = 1.2 √ P
d = 1.2 √ P 80 MHz to 800 MHz
d = 1.2 √ P 800 MHz to 2.5 GHz
3 V/m
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer
and is the recommended
separation distance in meters
(m). Field strengths from fixed RF
transmitters as determined by an
electromagnetic site survey.
1-800-MEDLINE (633-5463) | medline.com 25
Electromagnetic
environment – guidance