Permanently Linking the Transmitter to an Auxiliary
Display
Please contact your Stryker rep in order to permanently link the display(s) with
a specific transmitter, as deemed appropriate. A permanent link may be desirable
in certain cases. Once the display(s)/transmitter set has been permanently linked,
the set will automatically link when powered on. Display(s)/transmitter sets that
have been set to permanent link mode should be kept in the same operating
room.
To return the display(s)/transmitter set to default linking mode please follow the
instructions below using a blue WiSe(TM) Linking Token.
1. Power on the display(s) and transmitter.
2. insert the blue token into the transmitter and remove when the token LED
turns green.
3. insert the blue token into each display, one at a time, and remove the token
when the token LED turns green.
4. Power off all devices.
5. Power on all devices.
Now the system can be linked as described in the default linking mode
procedure.