Password protection and basic program
Volume controlled
Start via water meter
▼▲
▼▲
▼▲
Volumen control /
Time priority
▼▲
26
Minimum interval = 1 liter, maximum interval = 9999 liters. The
analysis is carried out once the programmed water quantity has been
measured. The line and the measuring chamber are flushed prior to
the analysis (observe the programmed flush times).
Selecting the volume control
In
=>
Press "ENTER" to confirm the selection
(Es erscheint ein Sternchen " * " am Zeilenende)
Es erscheint das Menü
Enter the respective flow rate in liters
Press "ENTER" to confirm the entry
Selecting the type of water meter
Select the
=>
Select the water meter constant (liter/impulse)
(100 liters/impulse " * " is factory set)
Press "ENTER" to confirm the selection
The analysis is carried out once the programmed water quantity has
been measured. A regeneration message is always generated with
higher priority when the programmed interval time has been reached.
Selecting the volume control/time priority
The process is the same as for "Selecting the time control"
The process is the same as for "Selecting the volume control"
Press "ENTER" to confirm all the entries
, select
=>
menu
=>
=>
=>