Calibration Instructions - Weston 24-1001-W Manual Del Usuario

Ocultar thumbs Ver también para 24-1001-W:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

OPERATING INSTRUCTIONS
FIGURE 5
6. When weighing an item that
exceeds the scale's overload or max
capacity (150 Kg/330 LB), the display
will show an "Err" message (FIGURE
5). To eliminate the message, remove
the item from the scale and try again,
using an item within the confines of
the scale's capacity. If the scale's
max capacity is exceeded for an extended period of time, there is a risk of
permanent damage to the scale. Promptly remove the over-capacity item
when you see the scale displays the "Err" message.
NOTE: To accurately weigh an item, keep scale on level and stable surface,
place the item in the middle of the scale platform. This will ensure the most
accurate weight reading.

CALIBRATION INSTRUCTIONS

1. To calibrate the scale, press the "MODE" button for 3-5 seconds. The
display will show random digits, "CAL" and then "50.00 Kg" (FIGURE 6). Place
a 50 Kg/110 LB weight in the middle of the scale platform.
FIGURE 6
2. Press the "MODE" button again. The display will show "100.00 Kg" (FIGURE
7). Place a 100 Kg/220 LB" in the middle of the scale platform.
3. Press the "MODE" button again. The display will show "150 Kg" (FIGURE 8).
Place a 150 Kg/ LB" weight in the middle of the scale platform.
4. Press the "MODE" button again. The display will show "----" (FIGURE 9). This
indicates that the calibration is finished and the scale will display the time/clock.
FIGURE 7
FIGURE 8
FIGURE 9
Note: If the scale is not stable during the calibration, the display will only show
the calibration weight and the scale will then automatically power off after two
minutes and will display the time/clock.
7

Publicidad

Tabla de contenido
loading

Tabla de contenido