Operating Instructions
VistaScan Plus Image Plates
The following information supplements the installation and operating instructions of the unit.
As a general rule, the installation and operating instructions for the device must also be observed.
These instructions contain important information including safety instructions and information on
setup, electrical connections, commissioning, the disinfection process, the cleaning process, etc.
1

Intended purpose

The reusable image plates are used to store intraoral or extraoral x-ray images in dental applications.
2

Intended use

The image plates can be reused multiple times after being erased.
Use the image plates solely in combination with image plate scanners manufactured by or branded with
Dürr Dental and light protection covers manufactured by or branded with Dürr Dental.
Images plates should be read out immediately, in order to avoid loss of image quality.
For image plate preparation refer to installation and operating instructions of the image plate scanners ma-
nufactured by or branded with Dürr Dental.
3
Usage

3.1 Correct use of image plates

Detailed information about the compatibility of various types of image plate with various types of
image plate scanner model by Dürr Dental is available in the Downloads area at www.duerrdental.
com (document-no.: 2130100177).
WARNING
Risk of cross contamination when not using the light protection cover or when using the
light protection cover more than once
Do not use an image plate without a light protection cover.
i
Do not use the light protection cover more than once (disposable item).
i
CAUTION
The image data on the image plate is not permanent.
The image data is altered by light, natural X-ray radiation and scattered X-ray radiation. This will
lead to a reduction in diagnostic information and clarity.
Read the image data within 30 minutes of exposure.
i
Never handle exposed image plates without the light protection cover.
i
Do not subject an exposed image plate to X-ray radiation before or after the scanning process.
i
Do not X-ray during the scanning process if the unit is in the same room as the X-ray tube.