EGK 2-19 (+)
5.1.1 Overview of the menu items
Display of current temperature and operating state
____
Display
____
Displayed is the block temperature with a resolution of 0.5 °C / 0.75 °F. By pressing the Enter button brings the
Current temperature
display to the Main Menu. The unit of temperature is adjustable in the menu Global Settings (Celsius or Fahrenheit).
Main Menu
Display
Cooler display
Top Settings
Global Settings
Exit
Exit Main Menu
6 Maintenance
Always observe the applicable safety- and operating regula-
tions when performing any type of maintenance. Please refer
to the original operator's manual on the included CD or online
at www.buehler-technologies.com for maintenance informa-
tion.
7 Service and Repair
Please refer to the original operator's manual on the included
CD or online at www.buehler-technologies.com for a detailed
description of the unit including information on
troubleshooting and repair.
10
Bühler Technologies GmbH
____
____
Submenu
____
____
E219
Set temperature
____
Alarm hysteresis
____
Alarm hysteresis
____
Moisture detector calibration
____
Moisture detector sensitivity
____
Exit Submenu
____
toP
____
E
Setting range
____
Parameter
____
Cooler temperature
____
___
Parameter
Upper alarm
____
___
Parameter
Lower alarm
____
Parameter
rSt
____
Parameter
HiLo
____
Parameter
C
C - F
____
adjust value
store value
wait 5s: no storage
Adjust the set point:
2°C...20°C / 35.6°F...68°F
Set the upper alarm threshold above
setpoint: tEMP 1°C...7°C / 1.8°F...12.6°F
Set the lower alarm threshold below
setpoint: tEMP -1°C...-3°C / -1.8...-5.4F
If moisture detector is installed:
calibration key
If moisture detector is installed:
adjust sensitivity: High, Low
Set unit Celsius or Fahrenheit
BX450010 ◦ 10/2018