Troubleshooting list

Malfunction
The blade set cannot be
fixed on hair clipper.
Stops during cutting
Upper blade vibrates.
Hair clipper doesn't charge. Wrong mains power supply
Blades do not move
Cause
The blade notch doesn't fit the
shaft properly.
Low power
Blade set is blunt.
Mains power supply not plugged in
completely.
Battery is comsumed/broken.
Low power
Battery is comsumed/broken.
Motor is broken.
Blade is not fixed properly.
Remedy
Pull small cutter back to center.
Connect hair clipper to mains power sup-
ply.
Replace blade set.
Contact distributor.
Replace mains power supply.
Plug in mains power supply completely
into DC socket 4 of the hair clipper.
Contact distributor.
Connect hair clipper to mains power sup-
ply.
Contact distributor.
Contact distributor.
Disassemble blade set and lock the blade
set again into position.