greenbite
Bite registration material, silicone based, addition curing
fast setting – extreme high final hardness
1. Mixing and Dosing
Extrude material from the cartridge with Automix2-system gun (fig. 1). Place
cartridge into mixing gun and remove closure cap by rotation. Eject a small
amount of material (bleeding) until equal flow is being extruded from both
orifices (fig. 2). Insert mixing cannula (green) into guide grooves on cartridge and
cannula and rotate anti-clockwise to a stop (fig. 3). Store cartridge with mixing
cannula attached for subsequent use.
Fig. 1
2. Instruction for use
Extrude material from cartridge with mixing cannula or with attached contouring
tip onto teeth (fig. 4). At terminal occlusion application from vestibular is possible. Let
registration harden in terminal occlusion and remove from the mouth after 1 min.
3. Disinfection
After removal from the mouth rinse registrate under lukewarm water.
Fig. 2
Disinfection for 15 min. in 2 % glutaraldehyde.
Important working hints
Remove any retraction astringents or mouth wash solutions prior to the
registration – could affect setting reaction.
Do not intermix with condensation curing silicones.
Avoid staining of clothing. Hardened impression materials cannot be removed.
Remove all residues of registration material from the patient's mouth.
For reducing the pressure needed during injection, we recommend using the
Fig. 3
green mixing cannula.
Further information:
Silicone based materials are proven a million times. On condition of a proper
application, undesired effects are not to be expected. However, reactions of the
immune system like allergies, irritations, cannot be absolutely excluded. In case of
doubt, we recommend to make an allergy test before the application of the
Fig. 4
material.
For use by trained specialists.
30 sec.
Working time
1 min. 30 sec.
Setting time
GmbH & Co. KG
Carl-Zeiss-Str. 4 · 76275 Ettlingen/Germany
Telefon: 0 72 43/510-0 · Fax: 0 72 43/510-100
www.detax.de · post@detax.de
apple
1 min.
Time in the mouth
GB
Instructions for use
Indications
for use:
Bite registration with
crown and bridge
impressions and for
orthodontic indications,
key for intraoral
registration
Technical
Data:
DIN 13903
Mixed volume:
50 ml (cartridge)
Mixing ratio: 1:1
Colour code:
base: green
catalyst: yellow
Mixing time:
omitted (Automix2-
system)
Working time:
approx. 30 sec.*
Setting time:
approx. 1 min. 30 sec.*
Time in mouth:
approx. 1 min.
Hardness Shore D: 45
Linear dimensional
change: < 0.2 %
Application:
At 23 °C ± 2 °C / 73 °F ±
4 °F, 50 ± 5% rel.
humidity
Storage:
* from beginning of mixing at
23 °C ± 2 °C / 73 °F ± 4 °F, 50 ± 5
% rel. humidity. Increased tem-
peratures accelerate, decreased
temperatures retard a. m.
times.
Ordering
information:
greenbite apple
Standard packing
2 x 50 ml cartridges
12 mixing cannulas, green
12 contouring tips
for mixing cannulas green
Ecopacking
8 x 50 ml cartridges
48 mixing cannulas, green
48 contouring tips
for mixing cannulas green
Mixing cannulas
green, 48 pcs.
Contouring tips
for mixing cannulas green
12 pcs.
milkbite
Standard packing
2 x 50 ml cartridges
12 mixing cannulas, pink
12 contouring tips
greenbite colour
Standard packing
2 x 50 ml cartridges
12 mixing cannulas, green
12 contouring tips
glassbite
Standard packing
1 x 50 ml cartridge
4 mixing cannulas, green
4 contouring tips
Made in
Germany
pixelbite
®
Standard packing
2 x 50 ml cartridges
12 mixing cannulas, pink
02/2012
12 contouring tips
0 2 6 4 1
0 2 6 4 7
0 2 7 0 5
0 2 6 4 8
0 3 3 8 0
0 2 2 0 7
0 2 5 7 1