Tightening Torques - WIKA TC82 Información Adicional

Ocultar thumbs Ver también para TC82:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

3. Commissioning, operation

3.2 Tightening torques

EN
Only ever screw in, or unscrew, the instrument via the spanner flats and to the
prescribed torque using an appropriate tool.
The correct torque depends on the dimensions of the connection thread and the sealing
used (form/material).
When screwing in or unscrewing the instrument, do not use the connection head as
contact surface.
When screwing in the instrument, please observe that the threads are not skewed.
Assure that the purge gas connection is tighten and no gas can escape, so that there is
no carry-through between zones.
8
Connection head, selectable (example)
Tightening torques between connection head and neck tube
Thread
1/2 NPT
3/4 NPT
M20 x 1.5 with counter nut
M24 x 1.5 with counter nut
Tightening torques between purge connection and purge piping
Thread
1/4 NPT
1) Turns from finger tight (T.F.F.T)
WIKA additional information TC82, TC83, flameproof enclosure (Ex d)
Tightening torques in Nm
Connection head material
Aluminium
Stainless steel
T.F.F.T 2 - 3
T.F.F.T 2 - 3
1)
T.F.F.T 2 - 3
T.F.F.T 2 - 3
1)
23
25
27
30
Tightening torques in Nm
Connection head material
Aluminium
Stainless steel
T.F.F.T 2 - 3
T.F.F.T 2 - 3
1)
1)
1)
1)

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Tc83Tc82-fTc83-fTc83-m

Tabla de contenido