3.4/ OVERLOAD INDICATION

Detection of overload of the range of the clamp is indicated by the red light
. This
indicator flashes for a peak current greater than 60 A on the 40 A range (10 mV/A) or
600 A on the 400 A range (1 mV/A).

3.5/ AUTO OFF

The clamp has an Auto Off feature which switches off 10 minutes after the clamp has been
switched on, without the controls being used.
When the clamp is switched off by this automatic function, the switch
must first be set to the
OFF position before being switched on again.
This function can be overridden by the user when switching on. Simply press the auto zero
button
at the same time as setting the switch
from the OFF position to the 1 mV/A or
10 mV/A position. If the green indicator
flashes whilst the zero reset button is being pressed,
this indicates that the auto off function has been inhibited.

4/ SPECIFICATIONS

Ranges
input/output ratio
measurement extent
AC rms
peak max
DC
40 A
10 mV/A
0.2...40 A
0.2...60 A
0.4...60 A
400 A
1 mV/A
0.5...400 A
0.5...600 A
0.5...600 A
4.1/ RÉFERENCE CONDITIONS
- Temperature: 18...28°C
- Humidity rate: 20...75% RH
- Battery voltage: 9 V ± 0.1 V
- Position of conductor: centred on the markings of the clamp
- Magnetic field: Earth's DC field
- Absence of external AC magnetic field
- Absence of electric field
- Measurement for a DC current or an AC sinusoidal current ≤ 65 Hz
- Impedance of the measurement instrument : ≥ 1 MΩ and ≤ 100 pF

4.2/ OPERATING CONDITIONS

The instrument must be used in the following conditions to satisfy the safety of the user and
the metrological performance:
- Use indoors
- Working altitude: ≤ 2000 m
- Transportation altitude: ≤12 000 m
- Environmental conditions: see graph below