Technical Data - Bowa 901-011 Instrucciones De Uso

Interruptor de pedal para usos médicos con cápsula de plástico
Ocultar thumbs Ver también para 901-011:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 26

10. Technical data

Standards:
DIN EN 60601-1:1990+A1+A2
DIN EN 60601-2-2:2007
DIN EN 60601-2-22:2005+A1:2007
DIN EN 60529:1991+A1
Classification: class IIb as per Medical Device Act / MDD 93/42/EEC
Housing and top (operating element):
unbreakable, self extinguishing thermoplast
Connection cable: firmly attached and poured control line
Protection: IP X8 as per IEC 60529
Switching element:
Reed contact
(1S, 2S, 1W)
Voltage: max. 25 V AC / 60 V DC
Supply voltage:
----
Current consumption:
max. 1 A
Breaking capacity:
max. 100 VA
Mechanical life: > 1 Mio. switching cycles
Approval: suitable for AP area
Storing conditions:
Ambient
conditions:
Page 5
Microswitch
(1PW)
----
max. 5 A
max. 1250 VA
Temperature
-20 °C to +70 °C
-10 °C to +60 °C
MN031-426-S2
IEC 60601-1:2007+A1+A2
IEC 60601-2-2:2010
IEC 60529
HALL sensor
(HS 0-10 V, 0(4)-20 mA)
15 V DC < U
max. 60 mA
----
Relative humidity
10 - 100%
10 - 100%
> 30 V DC
b
Air pressure
500 - 1060 hPa
800 - 1060 hPa
Medical foot switch