ENG
1.3.1. WATER DISTILLER. FRONT PANEL
Distilled
water
outlet
ON/OFF
switch
1.4. INSTALLATION
1. Connect the equipment to a 230 V power outlet using the power supply cable (110 V for
10DESA1041 and a three-phase connection for 10DESA0120)
2. Connect the gray hose to the water supply socket. The other end of the hose, with the filter
previously installed, will be connected to a ¾ "tap.
3. The silicone tube must be connected to the outlet for distilled water and the other end to a collecting
container whose level is below the outlet by about 25 cm.
DESA 0_ _ _
Refrigerante
Boiler
10
Steam exhaust hole
Water inlet
Silicone gasket
Level
Level switch
Water drain valve
Reset