ALARM (ERROR CODE)
CLOSE THE DOOR
STERILE ERROR
DOOR OPEN
PROCESS STOPPED
HEATING ERROR
POWER INTERRUPT
SERVICE
NO WATER
SERVICE HELP IS ALWAYS REQUIRED WHEN ANY OF THE ALARMS BELOW ARE DISPLAYED.
GETINGE PACS 300,
PT-100 ERROR,
PRESSURE SENSOR ERROR
UPDATE,
MANUAL MODE, POST TREATMENT
ERROR
SYMPTOM
PROCESS ABNORMALLY LONG
INCORRECT OPERATION, CHECKLIST
The table below lists the most common operating
faults that result in an alarm (error code) or that cer-
tain symptoms are experienced.
Check that usage was correct.
REASON/ACTION
Door has not been locked before the start button has been pressed.
The sterilizer may have been overloaded. Read the Technical data sec-
tion.
The door handle has been activated before the finished signal has been
displayed.
The stop button has been pressed by the operator.
Can be due to items caught in the door.
Check the fuses on the incoming supply
Time for the annual service. Contact service
Check the incoming water. Pressure and flow
Contact service
REASON
Can be due to:
no water on the incoming supply.
sterilizer overloaded. Read the Technical data section.
leakage.
OPERATION, CHECK