Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

1.3.

Operation

Leak Detection Sequence:
1) Press the start button.
a) The "Evac" diode comes on and the test object is evacuated.
Pre fill evacuating time is set and adjusted by a timer in Sensistor
ISH2000.
b) The "Fill" diode comes on and the "Evac" diode goes off.
Note: The "Evac" diode goes off shortly after the "Fill" diode
has come on.
Test object is now filled with tracer gas. Filling pressure is
controlled by the gas bottle regulator.
2) Search and locate leak(s).
3) Press stop button.
a) The "Evac" diode comes on and the test object is evacuated.
b) After test evacuation time is set by a timer in Sensistor
ISH2000.
c) Evacuation stops and Test object is refilled with air to
atmospheric pressure
INFICON – User's manual TGF10
EN
5

Publicidad

Tabla de contenido
loading

Tabla de contenido