General Specifications - Uline H-7095 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

FUNCTION
Continuity
Diode

GENERAL SPECIFICATIONS

SPECIFICATION
Display
Over-Range Indication
Conversion Rate
Maximum Voltage
Polarity Indication
Low Battery Indication
Auto Power OFF
Operating Temperature and Humidity
Storage Temperature and Humidity
Operating Altitude
Battery Power
Fuse Protection
Weight
Dimensions (W x H x D)
Safety Standards
EMC
PAGE 13 OF 39
SPECIFICATIONS CONTINUED
RESOLUTION
Continuity threshold is 10Ω ~ 100Ω.
For measurements < 10Ω the buzzer will sound.
0.1Ω
For measurements >400Ω (>600Ω ex355) the
symbol will display.
Open-circuit voltage is approx.: 3.2V.
1mV
Typical voltage of a silicon PN junction is in the
range of: 0.5 to 0.8V.
EX350 4000 count, MultiFunction LCD
OL
is displayed
Two to three updates per second
600VAC RMS or 600V DC maximum applied to any terminal
Automatic: positive implied, negative indicated (-)
is displayed.
After 15 minutes (can be disabled by holding the M [MODE] button while
turning the meter ON)
14 to 50ºF (-10~10ºC)
30 to 86ºF (0 to 30ºC); < 75% RH maximum
86 to 104ºF (30 to 40ºC); < 50% RH maximum
14 to 122ºF (-10 to 50ºC); < 80% RH maximum (with battery removed)
2,000 m (6,562')
2 AA batteries (battery life 130 hours typical with alkaline battery types)
10A terminal (F2): F 10A H 600V fast blow fuse (6.3x32) mm
mA, μA terminal (F1): FF 600mA H 600V fast blow fuse (6.3x32) mm
Including battery 407 g (14.36 oz)
7.1 x 3.4 x 2.2" (180 x 86 x 55.6 mm)
Complies with CSA STD C22.2 NO.60101-1, 61010-2-030, IEC61010-2-033,
CAT III 600V, Pollution Degree 2
EN61326 1
1-800-295-5510
uline.com
REMARKS
DESCRIPTION
OL
0621 IH-7095

Publicidad

Tabla de contenido
loading