Chapter 5
|
Preventive Maintenance
CAUTION
5-2
|
Isotemp

Electrical Power Cord

Ensure any replacement cord is properly rated.

Optical Disc

(SWB only)
Check the optical disc monthly and verify it is free of foreign material. If not, wipe the disc clean.
1. Remove the four screws, two on top and two on the side securing the top right panel.
2. Swing panel assembly open.
Do not operate the bath with the panel open.
3. Rotate disc, by grasping aluminum components of the eccentric only, one turn clockwise and
verify that disc does not rub against optical sensor. If required, use the 3/32" Allen to loosen two
set screws on the eccentric and reposition the disc.

Shaker Assembly

Check the shaker assembly for looseness annually. Contact Fisher Scientific if excessive
movement is noted.
Optical Disc
Set Screws
(located under the
disc)
Fisher Scientific