5.2 Fuse replacement
Before replacing the fuse, disconnect test leads and/or any connectors
from any circuit under test. To prevent damage or injury replace the fuse
only with specified ratings.
1. Set rotary switch to the OFF position.
2. Disconnect test leads and/or any connectors from the terminals.
3. Use a screwdriver to unlock the four screws on the rear cover.
4. Take out the rear cover from the meter.
5. Remove the fuse by gently prying one end loose, then sliding the fuse out of
its bracket.
6. Install the replacement fuses only with specified ratings: F 0.63A/1000V
Ø10.3x38 and F 10A/1000V Ø10.3x38
7. Rejoin the rear cover and tighten the screws.
5.3 Battery replacement
To avoid false readings, which could lead to possible electric shock or personal
injury, replace the battery as soon as the battery indicator "*" appears.
Before replacing the battery, disconnect test leads and/or any connectors from
any circuit under test, turn the meter off and remove test leads from the input
terminals.
1. Set rotary switch to the OFF position.
2. Disconnect test leads and/or any connectors from the terminals.
3. Use a screwdriver to unlock the two screws on the battery cover.
4. Take out the battery cover from the meter.
5. Remove the used battery.
6. Replace with one new 9V battery (6F22).
7. Rejoin the battery cover and tighten the screws.