Sterrad
1.
Clean and prepare the camera head and cable as recommended in this
user guide. Ensure the soaking cap is installed.
2.
If using a sterilization tray (optional), follow any additional instructions
provided with the tray. Use only trays that are compatible with Sterrad.
Not all sterilization trays are compatible with Sterrad systems.
Using an incompatible tray may result in incomplete device
sterilization. Consult the instructions that came with your
sterilization tray to determine which sterilization method is
compatible with your tray and devices. If a compatible tray
is not available, the devices can be double wrapped prior to
using the Sterrad system.
3.
Double wrap the camera head and cable (or tray) prior to sterilization.
4.
Sterilize the camera head and cable following the instructions of the
manufacturer, using the Sterrad 100S, NX™, or 100NX Sterilization
System. Select the standard cycle.
5.
Allow the camera head, cable, coupler, and scope to completely dry
before reassembly. Any moisture on the threads will cause the camera
and coupler windows to fog during use.
Storage
Never store the device in a non-ventilated, humid environment such as a
carrying case. This may present an infection control risk.

Using Sterile Drapes

Using sterile drapes will ensure maximum longevity of your 1488 HD
Camera Head. For best results, follow the instructions provided by the drape
manufacturer.

Disposal

This product contains electrical waste or electronic equipment.
It must not be disposed of as unsorted municipal waste and
must be collected separately in accordance with applicable
national or institutional related policies relating to obsolete
electronic equipment.
The 1488 HD must be disposed of according to local laws and hospital
practices.