WARNING
!
Make sure the power cord and plug are in good condition and the
•
equipment is not damaged.
The air filter cover protects the device in the event of accidental liquid
•
spillage onto the device. Ensure that the air filter and air filter cover are
fitted at all times.
3
Connect one end of the air tubing firmly onto the air outlet of the VPAP III ST-A
(D-2).
WARNING
!
Only ResMed air tubing should be used with your flow generator. A
different type of air tubing may alter the pressure you actually receive and
reduce the effectiveness of your treatment.
4
Assemble your mask system according to the mask user instructions.
5
Connect your mask system to the free end of the air tubing (D-3).
The VPAP III ST-A system is now assembled (D-4). To start treatment, see
"Starting Treatment" on page 17.
Attaching a Humidifier
WARNING
!
When using a humidifier, position it lower than you, and at the same level
or lower than the VPAP III ST-A.
Note: You must activate the humidifier option in the menus if you are using a
HumidAire or Passover humidifier.
Attaching a HumidAire 2i Humidifier
Please refer to the illustrations in section E of the illustration sheet.
The HumidAire 2i attaches to the front of the VPAP III ST-A to provide heated
humidification. No other accessories are required for its use. The VPAP III ST-A
automatically detects the presence of the HumidAire 2i. No menu changes are
required. Please refer to the HumidAire 2i User's Manual for details.
Attaching a HumidAire 2iC Humidifier
Please refer to the illustrations in section F of the illustration sheet.
The HumidAire 2iC attaches to the front of a VPAP III ST-A to provide passover
humidification. No other accessories are required for its use. Please refer to the
HumidAire 2iC User's Manual for details.
Attaching a HumidAire Humidifier
Please refer to the illustrations in section G of the illustration sheet.
Medium size (52 cm) air tubing is a necessary accessory for connecting the
VPAP III ST-A to the HumidAire humidifier.
12