Anton Paar DMA 35N Manual De Instrucciones página 19

Densímetro portátil
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

°BAUMÉ
Display of Baumé degrees at 60 °F. The measuring range is 0 to 100 °Baumé.
After storing the setting with a long push of the "store" key, the temperature
coefficient can be set.
Set the correct temperature coefficient (from 0 to 19x10
following procedure:
- A short push of the "store" key increases the digit.
- A long push stores the selected number.
- Then the instrument switches off automatically.
°PLATO
Display of Plato degrees at 20 °C derived from density at measuring temperature. The
measuring range is 0 to 100 °Plato.
%ALC/V, %ALC/W
Display of alcohol concentration of ethanol/water mixtures in % by volume or by
weight at 20 °C derived from density at measuring temperature. The measuring
range is 0 to 100 %v/v (%w/w).
Switch from % by volume (V) to % by weight (W) using a short push of the
"store" key.
PROOF
Display of Proof degrees at 60 °F derived from density at measuring temperature. The
measuring range is 0 to 200 °Proof.
CUSTOM
A custom-tailored polynomial function can be stored. Please contact your local Anton
Paar distributor.
PEr
Displays the period value of the measuring cell.
rES
Displays a value for the factory adjustment.
B46IB01
) according to the