1
Introduction
•
The DMA 35
portable density meter measures the density of liquids in g/cm
N
kg/m
3
according to the U-tube principle.
•
Samples are filled into the measuring cell using the built-in pipette-style pump or
a syringe.
•
A temperature sensor measures the sample temperature right at the measuring
cell. The temperature is displayed and can be used internally for automatic
temperature compensation of the density reading.
•
Once activated, automatic conversion of density into relative density, density at
reference temperature or concentration (H
°Plato, % alcohol, etc.) is performed immediately.
•
1024 measuring results including temperature can be stored in the memory of
the DMA 35
. Results can be recalled individually, or transferred to a printer or
N
PC using an infrared/RS 232 interface port (option).
B46IB01
SO
, API number, °Brix, °Baumé,
2
4
EN
3
or
7