FAST WAY (TABLE FLOWRATE):
Read the pump model that you have on the label applied on the pump (see
figure on the side) and write it on the "Table Flowrate" parameter as follows:
EMPIRICAL WAY
- Install the pump on the system.
- Make her prime.
- Insert the suction tube into a measuring cup filled
with the ml of water to dose. It is advisable to define a quantity such that the pump doses for at least 40 sec / 1 minute.
For example:
-300ml for a 20 liters / h pump; -200ml for a 12litre / h pump; -100ml for a 6litre / h pump; -50ml for a 3litre / h pump; -30ml
for a 2litre / h pump.
Rev 1.1
In the Calibration Menu, select with cursor "Table
Flowrate" and press ENTER.
Read the flowrate on the pump label and use the ▼▲
buttons to select the value and press the ENTER button
whenever necessary to confirm the change.
If you want to esc programming and return to
measurement, press ESC for 3 seconds.
For safety reasons, the pump must be calibrated
with water
In the Calibration Menu, select with cursor "Calib.
Flowrate" and press ENTER.
With the ▼▲ buttons choose the value of the quantity
of product to aspirate and press the ENTER button
whenever necessary to confirm the change.
Press ENTER to start dosing.
As soon as the container is empty, press ENTER to
stop the pump dosing.
The pump calculates the flowrate and writes it. Press
ENTER to confirm it.
If the display shows "Flowrate Error" message repeat
the calibration by increasing or decreasing the quantity
of product.
If you want to esc programming and return to
measurement, press ESC for 3 seconds.
Cod. 00.009.123
23