environmental Conditions

Indoor use only.
Recommended calibration interval: One year
calibration cycle
Operating temperature: 0°C ~ 30°C (≤80% RH)
30°C ~ 40°C (≤75% RH)
40°C ~ 50°C (≤45%RH)
Storage temperature: -20 to +60°C, 0 to 80% RH
(batteries not fitted).
Operating altitude: 2000m (6562 ft)
Conductor Size: 40mm diameter.
Shock vibration: No shock, no vibration to the jaw tips.
temperature coefficient: 0.2 x (Specified accuracy) /
°C, < 18°C, > 28°C .
eMC: EN 61326-1
Overvoltage category: IEC 61010-1 300V CAT III.
Pollution degree 2
Measurement
Category
I
II
III
IV
12
application
Measurement Category I is for
measurements performed on circuits
not directly connected to the mains.
Examples include: Measurements on
battery powered equipment and specially
protected (internal) mains-derived circuits.
Measurement Category II is for
measurements on circuits directly
connected to the low voltage installation.
Examples include: Household appliances,
portable tools and similar equipment.
Measurement Category III is for
measurements performed in the
building installation. Examples include:
measurements on distribution boards,
junction boxes, socket-outlets and wiring
and cables in the fixed installation.
Measurement Category IV for
measurements performed at the source
of the low-voltage installation Examples
include: Measurements on primary
overcurrent protection devices and
electricity meters.