WARNING: Use caution when performing the following step. The
aspiration cannula of some tips is extremely delicate and handling
may damage the tip.
5. With the system's vacuum set at greater than 400mmHg, simulate
an occlusion of the aspiration port by blocking the port at the
tip with a gloved hand. Activate the vacuum and verify that the
vacuum level climbs to within 50 mmHg of the system setting.
Check for leaks or bubbles in the aspiration line while the vacuum
is active.
CAUTION: Do not use the handle or tip if there are any problems.
Proceed to the troubleshooting section or contact MicroSurgical
Technology.
CLEANING
CAUTION: Improper or inadequate cleaning may result in
particulate matter adhering to the instrument and exfoliation into
the surgical field. In addition, the function and life expectancy
may become compromised.
The recommended cleaning process for the Intertip and Intertip-T™
system is as follows:
1. DURING THE PROCEDURE
Immediately after use, soak I/A tips and handles in distilled water.
2. POST PROCEDURE CLEANING
Clean the devices thoroughly per following procedure:
1. Flush both lumens of the handle (tip attached, sleeve removed)
with 10cc of deionized water 3 times each.
2. Rinse all the devices in fresh deionized water.
3. Remove the tip, if applicable.
4. Sonicate the devices in fresh deionized water for 3 minutes.
5. Re-attach the tip, if applicable.
6. Flush both lumens of the handle with tip with 10 cc of air 3 times
ea.
7. Dry the device with clean lint-free cloth
8. Visually inspect the devices for cleanliness. Repeat cleaning steps
as needed to remove visible soil.
NOTE: To avoid damage due to incidental contact, the Intertip and
Intertip-T™ tips and handle should be kept apart and/or protected
from other instruments during cleaning and sterilization.
7