Descargar Imprimir esta página

Cleaning And Caring For Your Appliance; Guide For Corrective Maintenance - Terraillon NASAL CARE SYSTEM 20 Manual De Instrucciones

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 32
Notice_Mouche_bebe Terraillon.qxd
Compact electric nasal aspirator/irrigator with wrist attachment
➛ Whenever you do not intend to use the
appliance for a long period of time, remove
the batteries from the unit to avoid them
damaging the appliance in case of leakage.
VI. CLEANING AND CARING
FOR YOUR APPLIANCE :
A. Cleaning the irrigation system
1. Switch off the appliance and detach the
nasal irrigation system from the seal/outlet
orifice on the main unit.
2. Unscrew the nasal irrigation system tank/
bowl to open.
3. Remove remaining saline solution from the
tank/bowl and the nasal irrigation system
spray nozzle/cannula. Rinse with purified
water.
4. Caution :
a. Make sure the main unit does not get
sprayed or splashed with or immersed in
liquid.
b. Do not remove the irrigation and mist
spray nozzle, or the irrigation system can-
nula/cover as this may cause the appliance
to malfunction.
c. Remove the remaining purified water from
the irrigation system using paper tissues
then leave to dry for at least two hours.
d. Do not use boiling water to sterilize the
irrigation system.
B. Cleaning the vacuum extraction system
1. Switch off the appliance and detach the
nasal vacuum extraction device/system
from the main unit's seal/outlet orifice.
2. Unscrew the vacuum extraction system
tank/bowl to open.
3. Remove the remaining nasal secretions and
mucus in the tank/bowl, the cannula and
the nasal vacuum extraction system using
the cleaning brush and purified water.
4. After having rinsed the nasal vacuum extrac-
tion system with purified water, sterilize it with
boiling water.
14
19/02/08
E N
NASAL ASPIRATOR & IRRIGATOR
VII. GUIDE FOR CORRECTIVE
A. The light emitting diode (LED) is flashing
B. The irrigation system is malfunctioning
16:40
Page 14
5. Caution:
a. Make sure the main unit does not get
sprayed or splashed with or immersed in
liquid.
b. Remove the remaining purified water from
the irrigation system using paper tissues
then leave to dry for at least two hours.
MAINTENANCE :
1. The light emitting diode (LED) will begin
flashing when battery power is low.
2. Remark:
Change the batteries.
1. Check that the bowl in the nasal irrigation
system contains sufficient saline solution.
2. Check that the irrigation and mist spray
nozzle is not obstructed.
3. Check that irrigation and mist spray system
has been correctly connected to the main
unit's seal/outlet orifice.
4. Make sure you have sufficient battery
power.
5. Remarks:
a. Whenever the irrigation and mist spray
nozzle is blocked, use that part on the
appliance and plug in to reverse the air-
flow in the saline solution circuit to
unblock the blocked part.
b. Whenever the irrigation and mist spray
nozzle is blocked, use the cleaning brush
to clean the nozzle and tube.
c. Whenever the irrigation and mist spray
nozzle is blocked, blow into each canal
with your mouth to clean the nozzle.
d. Change the batteries.
C. The vacuum extraction system is
malfunctioning
1. Check that the vacuum extraction system
is perfectly airtight.
2. Check that the vacuum extraction system
is not obstructed with nasal secretions
and mucus.

Publicidad

loading