Ugin ARTIS Modo De Empleo página 29

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 76
5.ARROW KEY - to right
During programming, press this key to switch from one parameter to the next.
During a firing cycle, press this key to cancel the ongoing phase and move on to the next.
6. ARROW KEY - upward
When the furnace is idle, press this key to change the program number.
During programming, press this key to change the value of the flashing parameter.
During a firing cycle, press this key to cancel the ongoing phase and move on to the next.
7. ARROW KEY - downward
When the furnace is idle, press this key to change the program number.
During programming, press this key to change the value of the flashing parameter.
During firing, press this key to cancel the cycle and switch to IDLE.
8. ARROW KEY - to left
During programming, press this key to return to the preceding parameter.
9. PROGRAMMING KEY
Use this key to initiate the programming mode. After pressing
firing parameters.
Once the program is entered, press
10. VACUUM INDICATOR
This indicator shows that the pump is in operation.
The pump continues to operate until the programmed vacuum level has been reached.
It is normal for the pump to stop and start several times during the firing cycle.
11. HEATING INDICATOR
This indicator shows that the muffle is heating.
As the temperature rises, it goes on and off regularly as the muffle heats up at the programmed
rate.
again to "confirm" the program settings.
27
, you can enter or modify the
ARTIS

Publicidad

Tabla de contenido
loading

Tabla de contenido