TEMPERATURE CALIBRATION
This setting, made during factory calibration, should give complete satisfaction while the muffle is
new (SET VALUE = 5).
Press on
The display screen will show.
• If the result indicates on overly high firing temperature:
increase the setting with
Press on
• If the result indicates an inadequate firing temperature:
decrease the setting with
Press on
You can check the accuracy of the temperature setting by running a silver test. Place the test piece
at least 15 mm above the surface of the firing platform, on an insulating base.
- Once the cycle is complete, the whole of the silver wire should be sagging.
- If the wire has entirely melted, there has been overheating; the temperature should be lowered by
5° C.
- If the wire has not melted, increase the setting by 5° C.
- Re-start the operation until a good result is achieved
CALIBRATION : 5
PENTE V:
button:
key to confirm.
key to confirm.
Then program the parameters as follows
- Pre-drying temperature
- Platform rise time
- Internal pre-drying
- Vacuum
- Vacuum level
- Hold
- Heat-rise rate
- Final temperature
- Hold time
- Internal cooling time
- Platform lowering time
50° C/min
6= 1 point or - 5° C
7= 2 points or - 10° C
button :
4= 1 point or + 5° C
3= 2 points or + 10° C
= 600° C
= 0 min
= 0 min
= 0° C
= 0%
= 0
= 50° C/min
= 960° C
= 1 min 30
= 0
= 0