8. Cleaning the cellular glass optic rod
(M500L, M600L, M500KL, M600KL, M500SL, M600SL, M500WL, M600WL, M500WLED, M600WLED, M500BL, M600BL, M500BLED, M600BLED)
Wipe clean the cellular glass optic rod entry point (Fig. 7) and exit point (Fig. 8) with an alcohol-immersed cotton swab. Remove all debris and oil.
Cotton Swab
Cellular glass optic rod
entry point
Fig. 7
CAUTION
Do not use a sharp tool to clean the cellular glass optic rod. It could damage the glass and reduce the light transmission. If illumination becomes
dim please contact your dealer.
8
Cotton Swab
Cellular glass optic rod
exit point
Fig. 8
9. Replacing the Turbine Cartridge
(1) Remove the turbine cartridge from the handpiece
1) Insert a bur into the chuck.
2) Locate the head cap wrench tool on the head cap then turn the
wrench counter-clockwise. Remove the cap.
3) Use the bur to gently lever the entire turbine cartridge out of the
head.
4) Remove debris from inside the empty head.
(2) Insert a new NSK turbine cartridge
1) Gently insert the new NSK turbine cartridge into the head section,
always making certain that the turbine cartridge locating pin is
aligned in the slot inside the head.
2) Use fingers to screw the head cap back into place and then use the
cap wrench to further tighten until the cap is securely in place.
Optional Cartridge : SX-MU03
Order No.P1001250
(Cartridge for Miniature head)
SX-SU03
Order No.P1002250
(Cartridge for Standard head)
Slot
Locating Pin
Head Cap Wrench
Head Cap
Cartridge
Fig. 9
9