Gossen MetraWatt SECULIFE ESPRO Manual De Instrucciones página 42

Analizador de aparatos eléctricos de cirugía
Tabla de contenido

Publicidad

SECULIFE ES
PRO
Parámetro CONFigure Subsystem
Ese grupo permite ajustar el display y las funciones operativas del analizador.
KEYWORD
CONFigure
:DISPlay
display_screen_number = 1-7
1 = One Parameter
2 = Two Parameters
3 = Three Parameters
4 = Four Parameters
5 = Five Parameters
6 = Measurement List Display (Non-editable)
7 = Quick Configuration Screen (Non-editable)
:SCReen
:AVERaging
:HOLD ON | OFF
:LOAD
:MODE
:VALue
:SETn
:NUMber
:Ln
INPut:
Atten:
RANGe:
NUMsamples:
MODE:
42
PARAMETER FORM
:SxZy S<display_screen_number> Z<zone_number> nn
< numeric_value >
FAST | SLOW | MEDium
TABle | ADJustable
< numeric_value >
< numeric_value >
< numeric_value >
< numeric_value >
0.1 | 1
100 | 1000 | AUTo
1024 | 2048 | 4096 | 8192 | 16384 | 54768
CONTinous | PULs
COMMENTS
nn = Parameter for selected Zone
0 = mV RMS
1 = mA RMS
2 = Watts RMS
3 = mV Peak
4 = mV Pk-Pk
5 = mV Pk / mV Pk-Pk
6 = mV Pk+
Range 1-7
1-5 = # display zones
6=Measurement List Display
7=Quick Config Screen
Table Mode: 1-12 para resistores
individuales, 13-22 para juegos de
resistores 1-10
Adjustable Mode: 0-6500.0
n = resistor a configurar
(1-10)
< numeric_value > = 16 bit
valor binario de resistores que
forman parte de juego 'n', siendo bit
0 = Load 1, Bit 1 = load 2...
Bit 11 = Load 12
cada bit determina si la carga
forma parte o no del juego.
Bit = 1 carga incluida
Bit = 0 carga excluida.
juego de valores numéricos
n = carga a configurar
< numeric_value > = valor actual de
carga 'n';
0-6500,0 óhmios
7 = Creast Factor
8 = Time-Pulse On
9 = Time-Pulse Off
10 = Time-Total Cycle
11 = % Duty Cycle
12 = mV Pulse Cycle
13 = mA Pulse Cycle
14 = Watts Pulse Cycle
GMC-I Messtechnik GmbH

Publicidad

Tabla de contenido
loading

Tabla de contenido