Pro - Comandos De Comunicación - Gossen MetraWatt SECULIFE ESPRO Manual De Instrucciones

Analizador de aparatos eléctricos de cirugía
Tabla de contenido

Publicidad

SECULIFE ES
PRO
SECULIFE ES
Palabra
clave
Nodos
CONFigure
DISPlay
HOLD
LOAD
INPut
SYSTem
POWerup
CONtrast
KEY
VERsion?
44
- comandos de comunicación
PRO
Subnodos
Valores
x is the Screen # (1-5) and y is the Zone # (1-5).
nn=0 to 13:
0=mV RMS
1=mA RMS
2=Watts RMS
3=mV Peak
4=mV Pk-Pk
5=mV Pk / mV Pk-Pk
SxZy nn
6=mV Pk+
7=Crest Factor
8=Time-Pulse On
9=Time-Pulse Off
10=Time-Total Cycle
11=% Duty Cycle
12=mV Pulse Cycle
13=mA Pulse Cycle
14=Watts Pulse Cycle
Range: 1-7
1-5 = # display zones
SCReen
6=Measurement List Display
7=Quick Config Screen
AVERaging
FAST, SLOW, MEDium
ON,OFF
MODE
TABle, ADJustable
Table Mode: 1-12 for individual Resistors, 13-22 for resistor sets
VALue
1-10
Adjustable Mode: 0-6500.0 Ohms
n = Resistor set to configure, 1-10
XXXX = 16 bit Binary value of resistors to include in set 'n'
Where Bit 0 = Load 1, Bit 1 =Load 2... Bit 11 = Load 12
SETn xxxx
Each bit selects whether the load is included in the set.
Bit = 1 includes the load, Bit = 0 excludes the load.
Ejemplo: el valor de 9 se corresponde con carga 4 y 1, exluyendo
todas las demás cargas
NUMber
1-12 (cargas existentes en el sistema)
n = Load to configure
Ln xxxx
xxxx = Actual value of load 'n', 0-6500.0 ohms.
ATTen
0.1, 1 (donut attenuation)
Rango
100, 1000, AUTo
NUMsamples
1024, 2048, 4096, 8192, 16384, 32768
MODE
CONTinuous, PULsed
DEFaults, CUStom, SETCurrent
0-20
DUP, DDN, SUP, SDN, VUP, VDN, SETup, RETurn,
GSAVe, HOLD
[read only]