Agilent Technologies E3610A Manual Del Usuario página 59

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 39
to the full rated value of power supply. Input off the electronic
load.
b. Connect the unit to the ac power line through a variable
autotransformer which is set for low line voltage (104 Vac for
115 Vac).
c.
Turn the unit's power on and turn CURRENT control fully clock-wise.
d. Turn up output voltage to the full rated values as read on the
digital voltmeter. Then input on the electronic load.
e. Record the output voltage at the digital voltmeter.
f.
Adjust autotransformer to high line voltage (127 Vac for 115
Vac).
g. Record the output voltage again immediately. The readings' dif-
ference during the immediate change should not more than
0.01% of output voltage plus 2 mV.
Load Transient Recovery Time
Definition: This is the time for the output voltage to return to within a
specified bend around its voltage following a step change in load.
Test Parameters:
Measured Variable: Output Voltage Transients
Expected Results: Less than 50 usec (at 10 mV from base line)
Test Procedure:
a. Connect the test equipment as shown in Figure A-2, but
replace the DVM with the oscilloscope. Operate the electronic
load in constant current mode.
b. Turn the unit's power on and turn CURRENT control fully clock-
wise.
c. Turn up output voltage to the full rated value.
d. Set the electronic load to transient operation mode between
one half of unit's full rated value and unit's full rated value at a 1
KHz rate with 50% duty cycle.
Figure A-3. Load Transient Recovery Waveform
e. Set the oscilloscope for ac coupling, internal sync and lock on
either the positive or negative load transient.
f.
Adjust the oscilloscope to display transients as in Figure A-3.
g. Check that the pulse width of the transients at 10 mV from the
base line is no more than 50 usec as shown.
PARD(Ripple and Noise)
Definition: Periodic and random deviation(PARD) in the unit's output
ripple and noise combine to produce a residual ac voltage superim-
posed on the dc output voltage. Constant voltage PARD is specified
as the root-mean-square(rms) or peak-to-peak(pp) output voltage in a
frequency range of 20 Hz to 20 MHz.
PARD(RMS) Measurement
Test Parameter:
Measured Variable: Output Voltage(rms)
Expected Results: Less than 200 uV rms
Test Procedure:
a. Connect the test equipment as shown in Figure A-4.
b. Turn the unit's power on and turn CURRENT control fully clock-
wise.
c. Turn up output voltage to the full rated value. Check that the
unit's CV indicator remains lighted. Reduce VOLTAGE control if
not lighted.
d. Check that the rms noise voltage at the true rms voltmeter is no
more than 200 uV.
A-3