Options

The following options are available with this instrument.
Option
Description
0EM:
115 Vac ± 10%, 47-63 Hz Input
0E3:
230 Vac ± 10%, 47-63 Hz Input
0E9:
100 Vac ± 10%, 47-63 Hz Input
Instrument Identification
Agilent Technologies power supplies are identified by one
serial number. The letter "KR" designates Korea as the coun-
try of manufacture, the first digit indicates the year (1 = 91, 2
= 92, etc), the second two digits indicate the week, and the
last five digits of the serial number are a different sequential
number assigned to each power supply.
If a yellow Change Sheet is supplied with this manual, its pur-
pose is to explain any differences between your instrument
and the instrument described in this manual. The Change
Sheet may also contain information for correcting errors in the
manual.
INSTALLATION

Inspection

When you receive your power supply, inspect it for any obvi-
ous damage that may have occurred during shipment. If there
is damage, notify the carrier and the nearest Agilent Sales
Office immediately. Warranty information is printed on the
inside front cover of this manual. Save the shipping carton
and packing materials in case the supply has to be returned
to Agilent Technologies in the future. If you return the supply
for service, attach a tag identifying the owner and model num-
ber. Also include e brief description of the problem.
Figure 1: OUTLINE DIAGRAM
The "Turn-On Checkout Procedure" in this manual can be
used as an incoming inspection check to verify that the sup-
ply is operational. See the appendix for tests that verify the
supply's specifications.

Location And Cooling

Figure 1 shows the outline shape and dimensions of the unit.
It is shipped ready for bench operation after connection to an
ac power source. The supply is air cooled. Sufficient space
should be allotted so that a free flow of cooling air can reach
the rear of the instrument when it is in operation. It should be
used in an area where the ambient temperature does not
exceed 40 degrees C. The current derates 1% per degree C
between 40°C-55°C.

Input Power Requirements

Depending on the line voltage option ordered, the supply is
ready to be operated from one of the power sources listed in
the Specification Table. The input voltage range, and the input
current and power at high line voltage and full load is listed for
each option.

Power Cord

This instrument is equipped with a three conductor power
cable. The third conductor is the ground conductor and when
the cable is plugged into an appropriate receptacle, the
instrument is grounded. The offset pin on the power cable
three prong connector is the ground connection. In no event
should this instrument be operated without an adequate cabi-
net ground connection.
The power supply was shipped with a power cord for the type
of outlet used at your location. If the appropriate cord was not
included, contact your nearest Agilent Sales Office to obtain
the correct cord.