MULTI-PURPOSE WORKSTATION™ STABILIZER LP (SINGLE-SIDED MOUNT)
SETTING UP ARM (SINGLE-SIDED MOUNT)
1. Loosen rail clamp levers on table mount. Slide table mount onto table rail and tighten rail clamps levers to secure. Center Y adjustment.
7. Fit dovetailed mounting plate to mating stabilizer mounting plate on table mount. Secure with stabilizer quick connect knob.
FUNCTIONAL PREUSE CHECKS
PRIOR TO USE
Attach table mount at a suitable clasping point on table rails to ensure it remains securely in position.
n
Ensure stabilizer moves freely when locking knob is loosened and secures firmly when locking knob is tightened.
n
Perform table clearance check. Before indexing the table, make sure stabilizer and table mount do not interfere.
n
REPROCESSING
WARNING
Users of this product have an obligation and responsibility to provide the highest degree of infection control to patients, co-workers and
n
themselves. To avoid cross-contamination, follow infection control policies established by your facility.
See your system's user guide for reprocessing transducer between uses.
n
These procedures for reprocessing have been validated for effectiveness and compatibility. Equipment could be damaged or cross-contaminated due
n
to improper reprocessing.
Do not place in a mechanical washer.
n
Do not use ethylene oxide or autoclave to sterilize.
n

CLEANING AND DISINFECTION

1. Using a clean germicidal wipe with at least 55% isopropyl alcohol and 0.5% quaternary ammonium, such as Super Sani-Cloth® Germicidal wipe,
wipe surfaces to remove gross soil.
Ensure all crevices and grooves have come into contact with wipe.
n
2. Using a new, clean germicidal wipe with at least 55% isopropyl alcohol and 0.5% quaternary ammonium, such as Super Sani-Cloth® Germicidal
wipe, wipe surfaces until visibly wet.
3. Allow treated surface to remain wet for no more than two (2) minutes.
Use additional wipe(s) if needed to assure continuous two (2) minute wet contact time.
n
4. Once device has reached two (2) minute wet contact time, allow surface to air dry.
I. Stabilizer Quick Connect Knob
II. Knob Y
III. Rail Clamp
IV. Table Rail
V. Rail Clamp Levers
VI. Stabilizer Mounting Plate
VII. Hex Shaft Attachment
VIII. Arm Locking Knob