Plateau
P 1500
set
Adjust (Plateau)
P 1500
set
Ramp rate C/min or F/min
set
Adjust (Ramp rate)
set
SWITCh TEST InSTruCTIonS
Test
Test
05
Test
05
Test
Switch test - plateau Set point
To check the starting point (Plateau)
for the initial temperature of the switch
test. Press the Set button twice. The P
set point temperature will be displayed.
Values between 10 and 700°C can be set.
Thermostats must not be connected to
any other voltage sources during test.
SWITCh TEST - Adjust plateau Set point
To Adjust the Plateau temperature press
the Set button twice and continue to hold
until the LED shown illuminates. Then
using just the UP or DOWN ARROW adjust
the Plateau temperature value. Press the
Set button once more to move on to the
ramp rate.
SWITCh TEST - ramp rate °C/minute
or °f/minute
To check the Ramp rate for the switch
test press the Set button three times. (If
checking after adjusting the Plateau press
the Set button once). The r value will be
shown as here. Values can be set between
0.1 and 10.0°C/minute (or 0.1 and 18.0°F/
minute).
SWITCh TEST - Adjust ramp rate
To Adjust the Ramp rate press the Set
button three times and continue to hold
until the LED shown illuminates. Then
using just the UP or DOWN ARROW adjust
the ramp rate value. Press the Set button
once more to move on to the Switch Test
End point.