The instrument should be carried either holding the front and the rear of the instrument or using the
carry strap provided. Never move or carry the instrument when in use or connected to the mains
electricity supply.
1.
All Techne instruments are supplied with a mains power cable.
2.
Before connecting the instrument to the mains electricity supply, check the voltage against the
rating plate (located on the side of the unit). Please note that the unit must be earthed to
ensure proper electrical safety. Connect the mains cable to a suitable plug according to the
table below.
Connection
Live
Neutral
Earth
3.
UK ONLY: The fused plug supplied with the mains cable is fitted with a 10 Amp fuse to protect
the instrument and the user.
4.
Units showing 230V, 50/60Hz on the rating plate also operate between 210 and 260V; units
with 120V, 50/60Hz also operate between 100 and 130V and units marked with 100V will
operate between 90 and 110V. In all cases the heating rate will degrade by approximately 8%
at the extremes of the voltage range.
5.
Plug the mains cable into the socket on the side of the instrument.
6.
Place the unit on a suitable flat bench ensuring that the air inlet vents on the underside are
free from obstruction.
7.
Switch on the instrument:
USB connection
with cover
Mains switch
On/Off
Fuses
InSTALLATIon
220/240V, 50/60hz Supply
Brown
Blue
Green/yellow
I mains switch On
o mains switch Off
4
110V/120V Supply, 50/60hz
Black
White
Green
Power inlet