User Reset

Reset the ORP calibration to the default using the user reset function.
1. Press
to enter setup window. Scroll down by pressing
display shows User Reset and Factory Reset.
2. Press
to select User Reset. The display automatically shows No and Yes.
3. Scroll down by pressing
4. Press
to confirm either No or Yes. The display shows the user reset option with a
Key Info
Scroll
Accept
EXIT

Factory Reset

Reset the tester to the factory default using the factory reset function.
1. Press
to enter setup window. Scroll down by pressing the
display shows User Reset and Factory Reset.
2. Scroll down by pressing the
automatically shows No and yes.
3. Scroll down by pressing
4. Press
to confirm either No or Yes. The display shows the factory reset option with a
Key Info
Scroll
Accept
EXIT

Manual Temperature Setting

The MTS range is 0°C to 60.0°C (32.0°F to 140.0°F). User reset will set temperature to default value 25°C or 77°F.
1. Press
to select setup window. Scroll down by pressing
The display shows Set °C/°F and Temp Cal.
2. Scroll down by pressing
3. The lower display shows the current measured temperature reading based on the last set offset and the upper
display shows the current measured temperature reading based on factory default calibration.
4. Dip the tester into a solution of known temperature and allow time for the in built temperature sensor to stabilize.
5. Press
to adjust the temperature value or press the
value of the solution.
Note: To exit without confirming the calibration, press
Key Info
Scroll
Accept
EXIT
4
to toggle between No and Yes.
Temp Set
User Reset
Settings
Factory Reset
Reset
to toggle between the resets. Press
to toggle between No and Yes.
Temp Set
User Reset
Settings
Factory Reset
Reset
to toggle between Set °C/°F and Temp Cal. Press
Temp Set
Set °C/°F
Settings
Temp Cal
Reset
to select Reset. Press
No
Yes
to select Reset. Press
to select Fact Reset. The display
No
Yes
to select Temp Set. Press
to confrim the calibrated value as new temperature
.
Temp Cal
23.4°C
23.4°C
23.4°C
Elite ORP Tester Operating Instructions
to select reset. The
.
to select Reset. The
.
to select Temp set.
to select Temp Cal.