Backlight Settings
1. Press
to enter setup window. Press
2. Scroll down by pressing
shows ON and OFF.
3. Scroll down by pressing
conditions.
4. Press
to select the desired backlight option. The display shows the selected backlight option with a
Key Info
Scroll
Accept
EXIT
Temperature Settings
1. Press
to enter setup window. Scroll down by pressing
The display shows Set °C/°F and Temp Cal.
2. Press
to select Set °C/°F. The display automatically shows °C and °F.
3. Scroll down by pressing
4. Press
to select the desired temperature unit.
Key Info
Scroll
Accept
EXIT
Temperature Calibration
1. Press
to enter setup window. Scroll down by pressing
The display shows Set °C/°F and Temp Cal.
2. Scroll down by pressing
3. The lower display shows the current measured temperature reading based on the last set offset and the upper
display shows the current measured temperature reading based on factory default calibration.
4. Dip the tester into a solution of known temperature and allow time for the temperature sensor to stabilize.
5. Press
to adjust the temperature value, then press
Note: To exit this program without confirming the calibration, press
Key Info
Scroll
Accept
EXIT
Elite pH and pH Spear Tester Operating Instructions
to toggle between Buffer and Backlight. Press
to toggle between ON and OFF. Backlight ON increases readability in low light
Settings
Buffer
Temp Set
Backlight
Reset
to toggle between °C and °F.
Settings
Set °C/°F
Temp Set
Temp Cal
Reset
to toggle between Set °C/°F and Temp Cal. Press
Settings
Set °C/°F
Temp Set
Temp Cal
Reset
to select Settings. The display shows Buffer and Backlight.
Backlight
ON
OFF
to select Temp Set. Press
°C
°F
to highlight Temp Set. Press
to confirm the new temperature value of the solution.
Temp Cal
23.4 °C
23.4 °C
to select Backlight. The display
OFF
°C
to select Temp Cal.
.
23.4°C
.
to select Temp set.
to select Temp set.
3