Removing the cartridge
1. Push up the lever [A] at the back of the
dispensing gun and pull back the slide [B]
completely.
2. Lift up the plastic lever [C] and remove the
cartridge.
Recommended use
• In order to improve the adhesion of the pre-
impression material to metal or plastic
impression trays, a tray adhesive for
A-silicones (e.g. tray adhesive, DMG) should
be used.
• The Intraoral-Tips are to be used for direct,
intra-sulcular application around the
preparation.
• The final impression can be optically
registered using an appropriate scanner or
cast in conventional fashion using plaster.
Double mixing impression
The individual times for the materials used
are provided in the „Technical data" section.
Follow the corresponding instructions for use!
1. Load the impression tray with Honigum Pro-
Heavy Scan. When doing so, hold the
impression tray at a slight angle beneath the
mixing tip.
Note: To avoid air bubbles, the end of the mixing
tip should remain immersed in the material
throughout application.
2. Inject around the preparation with Honigum
Pro-Light Scan.
Note: Filling of the impression tray and injection
should be concluded at the same time, as the
higher temperature in the mouth causes the
injected material to set more quickly than the
material in the impression tray.
3. Position the filled impression tray slowly and
gently in the mouth within the working time
and form the preparation.