Service information

Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to maintain
a device in accordance with the instructions may invalidate the device's warranty. Furthermore, failure to
maintain a device may compromise the clinical condition or safety of users and/or their caregivers. Do not
perform service and maintenance while the user is in the product.
1st Day
Read the User Guide carefully before use and
store it in a safe place for future reference
Store the enclosed tools on/at the product
Daily
Use a dry cloth to clean the product
Visual inspection. Repair or replace any
damaged or worn-out parts
Ensure that all fasteners are in place and fitted
correctly
Check all fittings and buckles for signs of worn
out parts
Weekly
Check that all wheels are moving freely and
that all wheel locks/brakes are easily activated
Wipe off any residue and dirt from the product
using a cloth with warm water and a mild
detergent/soap and dry before use
Disinfect the product. please check the
disinfection section in this User Guide for more
information
Monthly
All nuts and bolts on this product must be
checked and tightened to avoid unnecessary
failures
Oil the swing-away parts. We recommend you
use a professional lubricant system
Annually
Inspect the frame for cracks or signs for worn-
out parts and carry out annual service. Never
use a product which appears to be substandard
or faulty
Washing
The cushion covers can be washed at 60°C using
a mild detergent in a washing machine. It is not
recommended to wash the foam.
The foam inserts can be washed at 40°C using a
mild detergent in a washing machine.
During and after washing the foam, do not bend,
squeeze or wring it as it may negatively affect the
shape
Warewashing machine
This product can be washed at 60° using a mild
detergent in 10 minutes in a washing machine
designed for medical devices. Use the machine's
drying function to dry the product. Clean the
accessories seperately.
support.R82.org
Before washing
Avoid degreasing of the piston rods, the actuators
should be retracted to minimum stroke and without
load before washing.
Disinfection
The product without accessories and cushion can
be disinfected with a 70% disinfectant IPA solution.
It is recommended to wipe off any residue and dirt
from the product, using a cloth with warm water and
a mild detergent/soap without chlorine and let it dry
before disinfection.
Service interval
A detailed inspection of the product must be made
every 12 months (every 6 months for heavy users)
and every time the product is reissued for use. The
inspection should be carried out by a technically
competent person who has been trained in the use
of the product.
The lifetime of this product in normal use is 5 years
if all maintenance and servicing is carried out in
accordance with the manufacturer's instructions
and demonstrably recorded.
Service Check List:
Check and adjust the following functions as per their
intended use. Repair or replace any damaged or
worn-out parts. Please check the:
Seat plane adjustment. The seat plane
is adjustable and stays in position after
adjustment.
Frame is adjustable in height
Adjustment options for the foot rest, it is
possible to lock the angle and fasten the
telescopic tubes in position
Seat back is adjustable and can be locked in
position
Nuts and bolts on the product are tightened
Oil the swing-away parts
Product for cracks or signs for worn-out parts
All labels on the product are intact