7 Release the button when all patches in the strip have been scanned, and you reach the white
space at the end of the strip.
If the light on the EFI ES-2000 flashes green, the strip measurement is successful.
If the light on the EFI ES-2000 is blue, the first pass of a two-pass measurement is successful,
and the EFI ES-2000 is ready for the second pass of the same strip.
If the light on the EFI ES-2000 flashes red and an error message is displayed, you must
measure the same strip again.
8 For a two-pass measurement, repeat steps
Be sure to measure the strip in the direction indicated by the blue light on the EFI ES-2000.
The blue light indicates the leading side of the instrument.
Safety warnings
Note the following:
• Do not spill liquids on the EFI ES-2000.
• Do not drop the EFI ES-2000 into liquids.
• Do not try to dismantle the EFI ES-2000 for any reason. Unauthorized dismantling
of the device voids all warranty claims.
• Use the EFI ES-2000 on a stable surface and do not expose it to sunlight and moisture.
Maintenance and cleaning
Clean the plastic housing of the EFI ES-2000 with a damp cloth.
Clean the white reference ceramic tile on the calibration cradle with isopropyl alcohol on
a soft, clean cloth.
T
O CLEAN THE MEASUREMENT TUBE ON THE
1 Turn the EFI ES-2000 upside down.
2 Turn the aperture cover clockwise and remove it.
Avoid touching the optics.
3 Slide the slider with the protection glass out of the aperture.
4 Clean the glass and the inside of the aperture with isopropyl alcohol on a soft clean cloth.
5 Reassemble the aperture with the protection glass slider. Make sure that the slider snaps
back into its original position.
6 Insert the aperture cover and turn it counterclockwise to lock it in place.
Transportation
Transport the EFI ES-2000 packed in its case.