EFI ES-6000 Guia De Instalacion página 4

Espectrofotómetro
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 23
Clear a paper jam
1 Turn off the ES-6000.
2 Try to pull the page out of the instrument.
3 If you cannot pull the page out, disconnect the cables from the ES-6000.
4 Insert a flat screwdriver in the slot at the center of the cover and turn it to open the cover.
5 Move the measuring head carriage all the way to the left by pushing the plastic housing of the motor. Touch only
the motor.
6 Turn the roller wheel until the page is free.
7 Close the cover, reconnect the cables, and turn on the ES-6000.
Clean the white calibration tile
Clean the white calibration tile once a year or when the software indicates that cleaning is needed.
1 Turn off the ES-6000.
2 Disconnect the cables from the ES-6000.
3 On the bottom of the ES-6000, press the latch of the white calibration tile holder and flip it out.
4 Using alcohol and a cotton swab or lint-free wipe, clean the white tile.
5 Replace the white calibration tile holder.
6 Reconnect the cables.
Clean the internal feeding mechanism
Clean the internal feeding mechanism as needed.
Do not clean any parts of the instrument other than the ones that are specified.
Do not use compressed air to blow dust out.
Do not get dust on the white calibration tile.
1 Turn off the ES-6000.
2 Disconnect the cables from the ES-6000.
3 Insert a flat screwdriver in the slot at the center of the cover and turn it to open the cover.
4 Move the measuring head carriage all the way to the left by pushing the plastic housing of the motor. Touch only
the motor.
5 As you manually turn the roller wheel, use water or alcohol and a lint-free wipe to clean the bottom media guide
assembly and the rubber rollers.
6 Close the cover and reconnect the cables.
4

Publicidad

Tabla de contenido
loading

Tabla de contenido