Error Guide - iLine microsystems microINR Instrucciones De Uso

Ocultar thumbs Ver también para microINR:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

6. ERROR GUIDE

Error
Probable Cause
Messages when preparing to test
E01
The Datamatrix could not be read.
E02
Expired Chip.
E03
The 80 second interval for applying the
sample has been exceeded.
E04
Chip inserted backwards.
E11
Faulty or incorrectly inserted Chip.
Messages related to the test
E05/
Inadequate coagulation of the sample during
E09
the test.
E10
The INR value of the control channel is
outside the defined range.
E14/
Error while processing the sample during the
E15/
test.
E17
E16
Inadequate coagulation of the sample during
the test. ATTENTION: Possible sample with
abnormally high clotting times.
E18
Inadequate sample handling or hematocrit
outside the defined range.
Other messages
E06
Failure while checking the electronic
components of the Meter.
E07
Temperature below the defined range.
E08
Low battery.
E12
Temperature above the defined range.
Possible Solution
Insert the same Chip again, ensuring correct insertion. If the problem persists, repeat the
test with a new Chip.
Verify the date of the Meter.
If the date is not correct, enter the current date and insert the same Chip again.
If the date is correct, repeat the test with a new batch of Chips after verifying their expiry date.
If the sample has not yet been applied, repeat the test with the same Chip.
Rotate the Chip and repeat the test.
Insert the same Chip again, ensuring complete insertion. If the problem persists, repeat the
test with a new Chip.
Repeat the test with a new Chip. Strictly follow instructions on obtaining and applying the
sample.
If the problem persists, repeat the test with a new box or batch of Chips.
Repeat the test with a new Chip.
If the problem persists, repeat the test with a new box or batch of Chips.
Repeat the test with a new Chip.
Strictly follow instructions for conducting the test.
If the problem persists, repeat the test with a new box or batch of Chips.
Repeat the test with a new Chip.
If error E16 is displayed again, use a different measurement method.
This error occurs mainly in patients with high clotting times.
Repeat the test with a new Chip. Strictly follow instructions on obtaining and applying the
sample. If error E18 is displayed again, use a different measurement method.
This error occurs mainly in patients with a hematocrit outside the defined range for the
microINR system (25%-55%).
If the problem persists, contact your local distributor.
Repeat the test in a warmer location.
Charge the device with the charger supplied by the manufacturer.
Repeat the test in a cooler location.
15

Publicidad

Tabla de contenido
loading

Productos relacionados para iLine microsystems microINR

Tabla de contenido