Environmental Conditions; Mechanical Characteristics; Compliance With International Standards; Electromagnetic Compatibility (Cem) - Chauvin Arnoux PA32ER Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

3.2.4. CHARGER
Charging current: 1A.

3.3. ENVIRONMENTAL CONDITIONS

Range of operation: from -20 to +45 °C, from 30 to 95%RH without condensation.
Range of storage: from -25 to +70 °C, from 10 to 90%RH without condensation.
Electric field: < 1V/m
Magnetic field: < 40A/m
Indoor use.
Altitude : < 2000 m
Degree of pollution: 3

3.4. MECHANICAL CHARACTERISTICS

Dimensions (L x D x H)
Lead
Weight
Inrush protection
Drop

3.5. COMPLIANCE WITH INTERNATIONAL STANDARDS

The instrument is compliant with IEC/EN 61010-1 or BS EN 61010-1 and IEC/EN 61010-031 or BS EN 61010-031 for a voltage
of 1000V in overvoltage category IV with degree of pollution 3.
The instruments are compliant with FCC Rules part 1 under number 83TD07571.

3.6. ELECTROMAGNETIC COMPATIBILITY (CEM)

The instrument is compliant with standard IEC/EN 61326-1 or BS EN 61326-1.
220 x 112 x 53 mm
1.50m, terminated by a specific 4-point connector.
approximately 930g.
IP54 when not in operation and IP50 in operation per IEC 60 529.
IK 08 per IEC 62262.
1 meter.