Powermeter EduTrainer
®
Powermeter EduTrainer
Measurement uncertainty
Reference conditions
Voltage
Neutral current
Power
Power factor
Frequency
Imbalance U, I
Harmonics
THD U, I
Active energy
Reactive energy
Measurement with fixed mains frequency
General
Imbalance
Harmonics
THD, TDD
Nominal frequency
Sampling rate
Approvals
Subject to change
Further technical data can be found in the original operating instructions of the device.
They can be found on the manufacturer's website under
34
technical data
®
According to IEC/EN 60688, ambient 15...30°C,
sinusoidal input (form factor 1.1107), no fixed sampling frequency, measuring time 200 ms
(10 periods at 50Hz, 12 periods at 60Hz)
Current: ± 0.1% 1) 2)
± 0.2% 1) (if calculated)
± 0.2% 1) 2)
± 0.2°
± 0.01 Hz
± 0.5%
± 0.5%
± 0.5%
Class 0.5S, EN 62053-22
Class 0.5S, EN 62053-24
± Basic uncertainty x (F config–F actual) [Hz] x 10
U ± 2% to ± 0.5 Hz
± 2% bis ± 0.5 Hz
± 3.0% to ± 0.5 Hz
42...50...58 Hz or 50.5...60...69.5 Hz, programmable
18 kHz
CE marking in accordance with EMC, RoHS and Low Voltage Directive
The device is equipped with internal miniature fuses.
In the event of a fault, the device can only be repaired by Festo Didactic.
WARNING
www.camillebauer.com
SINEAX AM3000 (2019-03)
8108196
© Festo Didactic