Controls And Indicators - pico Technology A3000 Serie Guia De Inicio Rapido

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 20
Quick Start Guide
Typical connections
To obtain the highest bandwidth and best step response from your probe, you must minimize the
inductance of the signal path from the probe to the device under test:
• Use the shortest possible probe tip (ideally the fixed and sprung probe tips supplied)
• Use the shortest, widest possible ground connection (ideally the wide ground blade supplied)
To probe two points with 0.1" (2.54 mm) spacing, use the joggle adaptor. Performance will be
degraded slightly:
j PicoScope 6000E Series oscilloscope
k Joggle adaptor with probe tips
l Device under test

Controls and indicators

The A3000 Series probe has a pushbutton (m) and LED indicator (n), shown below:
m Pushbutton:
Press to start and stop capture in PicoScope.
n
LED indicator: Green when probe is active.
DO350-1
j
m
Flashes blue when oscilloscope is communicating with probe.
Yellow when probe is inactive.
Copyright © 2021 Pico Technology Ltd. All rights reserved.
A3000 Series Active Probe
k
n
l
5

Publicidad

Tabla de contenido
loading

Tabla de contenido