EMC Tables
Guidance and manufacturer's declaration – electromagnetic emissions
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the
product should assure that it is used in such an environment.
Emissions
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000 -3-3
Guidance and manufacturer's declaration 211; electromagnetic immunity
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
Surge
±1 kV line(s) and neutral
IEC 61000-4-5
Voltage dips, short
<5 % U
interruptions and
(>95 % dip in U
voltage variations
for 0,5 cycle
on power supply
40 % U
input lines
(60 % dip in U
IEC 61000-4-11
for 5 cycles
70 % U
(30 % dip in U
for 25 cycles
<5 % U
(>95 % dip in U
for 5s
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level
T
25
Compliance
The product uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference
Group 1
in nearby electronic equipment.
The product is suitable for use in all establishments, including
Class B
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Not Applicable
Not Applicable
Compliance level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV line(s) and neutral
<5 % U
T
)
(>95 % dip in U
T
for 0,5 cycle
40 % U
T
)
(60 % dip in U
T
for 5 cycles
70 % U
T
)
(30 % dip in U
T
for 25 cycles)
<5 % U
T
)
(>95 % dip in U
T
Not applicable
Electromagnetic environment-- guidance
Electromagnetic environment guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
T
)
typical commercial or hospital environment.
T
If a dips or an interruption of mains power
occurs, the current of the product may be
T
)
dropped off from normal level, it may be
T
necessary to use uninterruptible power
supply or a battery.
T
)
T
T
) for 5s
T
Not applicable