Prompt Message Of Instrument; Appendix 1. Continuous Output Format Specification; Appendix 2. Printing Format Specification - B3C pesage TRANSERENITE ARX Manual De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

VII
PROMPT MESSAGE OF INSTRUMENT 
The instrument has extremely high stability and reliability, thus is not easy to have error in general situation. Once an error occurs,
please make clear the error first and observe whether the instrument still has error after power-on. Do not hurry to repair the scale
body or instrument. Repair the instrument according to error code of the instrument as possible.
Nº 
Symbol 
Analysis 
1
Unable to reset after startup
[-EEE]
[ EEE]
2
The weighed object is over full range for 9 days
[
]
3
The weighed object is below 0 for 5
[
]
4
Out of clearing range
[
]
[
]
5
Invalid operation
6
EEPROM checksum and error
[Err 03]
7
The calibration input weight is too small
[Err 0 ]
8
The weight in calibration is too light
[Err 06]
9
The scale is dynamic in scale
[Err 07]
10
Setting error of date and time
[Err 08]
11
Error of AD initialization
[Err 09]
12
In loading scale, it indicates to loading weight; Load weight according to requirements
[LOAd]
13
It has enters menu setting
[ Et P]
14
End of point and loading point calibration
[End
]
15
Including current display weight in accumulated
[Add-- ]
value
16
Accumulated weight overflows
[-O ER-]
17
Loading default value
[Ld--- ]
18
Printing
[Print]
USER'S MANUAL
Remark 
1. Determine it is no-load state in startup;
2. Make calibration again.
Reduce weight on weighing platform
DAYS
Press
to reset
Check whether the weighing platform has weight.
Remove weight.
Press
reprint factory value. Start up again. If the
information occurs again, return to factory for repair.
Please calibrate the scale again if the situation does not
occur;
Attention: this place is provided with all parameters of
instruments of the factory.
Input≥10 % weight of full range
Load≥10% weight of full range
Inspect the scale body
Set according to specification of date and time
If the error occurs after restarting, return it to the fac-
tory for repair
Pres [print] to continue setting.
Clear accumulated weight in time.
TRANSERENITE ARX
13

Publicidad

Tabla de contenido
loading

Tabla de contenido